LPC-301 laser particle counter is used to measure the size and quantity of particles within air of unit volume in clean environment; it can directly measure clean environments from clean class 100,0000 to 10. This counter has used semiconductor laser power and nixie tube with high-luminance, so it is small, light, highly accurate, easing-operating and microprocessor-controlled. It may store and print the measure results, the printed result will show the clean class of that point and the measure of the clean environment is convenient. The product is widely used in electronics, optics, chemistry, food, cosmetic, medical science, biological products, aviation, and so on. Pick up Parts Characteristics: high accuracy thermal printer inside, can be transferred to M3 small, light, easy-operating, convenient short self-purifying time with flowmeter itself |
Main Technical Parameter
model |
LPC-301 LED |
LPC-301 LED pile |
Dimensions |
245×286×126mm3(W × D ×H) |
|
weight |
4. 0kg |
4.5kg |
Maximum power dissipation |
44W |
10W |
Power supply |
AC 100-240V,50/60Hz |
AC(100-240V 50/60Hz) and |
Particle size passage |
0.3、0.5、0.7、1.0、2.0、5.0、7.0、10( μm) |
|
Sampling flow rate |
0.1CFM(2.83L/min) |
|
Measure period |
1min、2min、3min、 10min |
|
Self-purifying time |
≤10min |
|
print |
Thermal printer inside, can be printed into grain/m3 |
|
Warning class |
100、10000 |
|
Usable environmental condition |
temperature:10-30℃ |
|
humidity20%~75% |
||
Atmospheric pressure:86Pa-106kPa |
||
Maximum sampling concentration permitted |
35000 grain/L(the particle size of a particle shall not be over 0.5μm, |
|
Continuous work time |
8 hours |
≥8 hours |
Light source and life |
Semiconductor laser,life is over 30,000 hours |
|
Packing of data |
N/A |