LPC-301 LCD (0.1CFM)

2014-04-10 11:50 PV: 3214

LPC-301 laser particle counter is used to measure the size and quantity of particles within air of unit volume in clean environment, it can directly measure clean environments from clean class 100,0000 to 10. This counter has used semiconductor laser power and big liquid crystal screen, so it is small, light, highly accurate, easing-operating and microprocessor-controlled. It may store and print the measure results, the printed result will show the clean class of that point and the measure of the clean environment is convenient.

Besides, the computer shall be on long-distance control through proprietary software. The product is widely used in electronics, optics, chemistry, food, cosmetic, medical science, biological products, aviation, and so on. 

Pick up Parts

characteristics:high accuracy thermal 

printer inside, can be transferred to M3small,
light ,easy-operating, convenient
short self-purifying time
with flowmeter itself
LPC-301 Laser Particle Counter LCD

 

Main Technical Parameter

model

LPC-301 LCD

LPC-301 LCD pile

Dimensions

245×286×126mm3(W× D ×H)

weight

4. 0kg

4.5kg

Maximum power dissipation

44W

10W

Power supply

AC  100-240V50/60Hz

AC(100-240V 50/60Hz)and 
DC11.1 V 8500mA( Li- ion batteries)

Particle size passage

0.30.513510μmshow 6 bunches of particles at the same time and directly show particle concentrationunitgrain/m3)

Sampling flow rate

0.1CFM(2.83L/min)

Measure period

1-40min

Self-purifying time

10min

print

Thermal printer inside, can be printed into grain/m3

Warning class

100100010000

Usable environmental condition

temperature10-30

humidity20%~75

Atmospheric pressure86Pa-106kPa

Maximum sampling concentration permitted

35000 grain/Lthe particle size of a particle shall not be over 0.5μm,

acid ,alkali and such corrosive gas are not permitted in the sampling air.)

Continuous work time

8 hours

8 hours

Light source and life

Semiconductor laserlife is over 30,000 hours

Packing of data

256 groups



 